Emerson Process Management 402 Marine Instruments User Manual


 
Instruction Manual 402 and 402VP
LIQ_MAN_ABR_402_402VP January 2014
E
NDURANCE
Conductivity Sensors
For additional information, please visit our website at www.rosemountanalytical.com
CAUTION
Sensor/Process Application Compatibility
The wetted sensor materials may not be compatible with process composition and operating conditions.
Application compatibility is entirely the responsibility of the user.
CAUTION
The 402 and 402VP sensors are retractable. BEFORE RETRACTING THE SENSOR, be absolutely certain the process pressure
is less than 64 psig (542 kPa abs) and the process temperature is at a safe level!
Sensor Specifications
Specifications 402 402VP
Electrode -11, -12, -13 titanium -11, -12, -13 titanium
Insulator
-11, -12, -13 PEEK (glass lled) -11, -12, -13 PEEK (glass lled)
Other Wetted Materials 316 SST, EPDM, graphite, Neoprene 316 SST, EPDM, graphite, Neoprene
Temperature Range 32–212 °F (0–100 °C) 32–212 °F (0–100 °C)
Maximum Pressure 200 psig (1481 kPa abs) 200 psig (1481 kPa abs)
Maximum Retraction Pressure 64 psig (542 kPa abs) 64 psig (542 kPa abs)
Sensor Orientation
Keep ¼ in. (6 mm) clearance between electrodes and piping. The electrolytes must be completely submerged in the
process liquid, i.e., to the upper edge of the guard (item 10 in Figure 3).
Figure 1. Sensor Orientation
°
°
30° 30°
trapped air
trapped sludge
Figure 2. Recommended Installation
recommended
not
recommended
flow
flow
flow