2. SPECIFICATIONS AND AGENCY
APPROVALS
2.1. Specifications
Measuring principle UV absorption (Beer Lambert Law)
Ranges
User selectable to any full scale range from:
0-1000 PPM to 0-10,000 PPM
0-5 g/Nm
3
to 0-50 g/Nm
3
0-1 wt% to 0-5 wt%
Measurement Units PPM, wt%, g/Nm
3
Accuracy
1% of Full Scale
Precision/Repeatability
0.5% of Full Scale
Display Resolution 1 PPM, 0.001 wt%, 0.01 g/Nm
3
Response Time (95%) <45 seconds to 95%
Compensation
Pressure, Temperature (NTP = 273.15K, 760
mmHg)
Gas Inlet Pressure Range 11.0 – 16.0 psia
Sample Flow Rate 0.8 LPM
Temperature Range 5-45
o
C
Dimensions (H x W x D)
Rack Mount (3U Panel) - 5.22” (133 mm) x 19.0"
(483 mm) x 15.3" (388 mm)
With NEMA 4X enclosure – 16.01” (407 mm) x
15.77” (401 mm) x 8.02” (204 mm)
Weight
Rack Mount – 13.6 lbs (6.17 kg)
With NEMA 4X enclosure– 16.5 lbs (7.5 kg)
Power 100-240 VAC, 47-63 Hz, 74 W Max
Maximum Operating Altitude 2000 meters
Analog Output 0 – 5 V or 4-20 mA isolated output
Status Outputs
System OK, Invalid Reading, UV Lamp Fault,
Pneumatic Error
RS-232 (optional RS-485) 57.6 Kbaud, DB-9 connector