Braun HNS 11 Marine Sanitation System User Manual


 
Appendix A
Stimulation current display in the mA range
Patient current Measurement display Controller display
0.00 mA 0.00 mA 0.00 mA
0.05 mA 0.05 mA ± 0.02 0.05 mA ± 0.03
0.10 mA 0.10 mA ± 0.02 0.10 mA ± 0.03
0.20 mA 0.20 mA ± 0.02 0.20 mA ± 0.03
0.30 mA 0.30 mA ± 0.02 0.30 mA ± 0.03
0.40 mA 0.40 mA ± 0.02 0.40 mA ± 0.03
0.60 mA 0.60 mA ± 0.02 0.60 mA ± 0.03
0.80 mA 0.80 mA ± 0.02 0.80 mA ± 0.03
1.00 mA 1.00 mA ± 0.02 1.00 mA ± 0.03
Patient current Measurement display Controller display
0.00 mA 0.00 mA 0.00 mA
0.50 mA 0.50 mA ± 0.1 0.50 mA ± 0.15
1.00 mA 1.00 mA ± 0.1 1.00 mA ± 0.15
1.50 mA 1.50 mA ± 0.1 1.50 mA ± 0.15
2.00 mA 2.00 mA ± 0.1 2.00 mA ± 0.15
3.00 mA 3.00 mA ± 0.1 3.00 mA ± 0.15
4.00 mA 4.00 mA ± 0.1 4.00 mA ± 0.15
5.00 mA 5.00 mA ± 0.1 5.00 mA ± 0.15
24
Appendix B
Conversion table: Amplitude (mA) and pulse width (ms) to charge (nano-Coulomb, nC)
Amplitude Pulse width Charge Pulse width Charge Pulse width Charge
mA ms nC ms nC ms nC
0.1 0.1 10 0.3 30 1.0 100
0.2 0.1 20 0.3 60 1.0 200
0.3 0.1 30 0.3 90 1.0 300
0.4 0.1 40 0.3 120 1.0 400
0.5 0.1 50 0.3 150 1.0 500
0.6 0.1 60 0.3 180 1.0 600
0.7 0.1 70 0.3 210 1.0 700
0.8 0.1 80 0.3 240 1.0 800
0.9 0.1 90 0.3 270 1.0 900
1.0 0.1 100 0.3 300 1.0 1000
1.2 0.1 120 0.3 360 1.0 1200
1.6 0.1 160 0.3 480 1.0 1600
1.8 0.1 180 0.3 540 1.0 1800
2.0 0.1 200 0.3 600 1.0 2000
2.5 0.1 250 0.3 750 1.0 2500
3.0 0.1 300 0.3 900 1.0 3000
3.5 0.1 350 0.3 1050 1.0 3500
4.0 0.1 400 0.3 1200 1.0 4000
4.5 0.1 450 0.3 1350 1.0 4500
5.0 0.1 500 0.3 1500 1.0 5000