Stearns Recreational 6340 Life Jacket User Manual


 
15
2. Oral Inflation Leak Test - This test should be performed
every two months and at the start of each season. Inflate
buoyancy chamber using the oral inflation valve and leave to
rest for 16 hours. When buoyancy chamber remains firm and
full of air there is no leakage and chamber is operational. If
the chamber deflates during this leak test, it is no longer in
operational condition and should not be used. Some types of
leaks can be repaired by a certified Stearns Repair Facility,
while others are not repairable. If the PFD leaks, either return
to certified Stearns Repair Facility for evaluation and repair, or
discard the PFD.
3. Function Test of Oral Valve - With the PFD inflated, remove
the dust protection cap, and test the valve by pushing it down
using the tab provided on the cap (refer to Page 10). The
valve should depress easily, and when depressed, air should
escape. When released the valve should automatically return
to the closed position and reseal.
REARMING KIT(S)
Use replacement kit no. 0925. See Rearming Instructions on
pages 11 and 12.
CARE AND CLEANING INSTRUCTIONS
Drying – If the device is wet, place on a coat hanger to allow
thorough drying. Do not place in direct sunlight to dry. Do not use
any heat source or place on radiator.
Cleaning – Damp wipe only with mild detergent recommended for
synthetic fabrics. Remove grease and oil immediately. After every
use rinse well with fresh water. Read instructions provided on the
device.
DO NOT MACHINE WASH
DO NOT MACHINE DRY
DO NOT DRY CLEAN
DO NOT USE SOLVENT BASED CLEANERS
DO NOT BLEACH
Revised: 2006-05-08