Teledyne HFC-E-202 Marine Sanitation System User Manual


 
HFM-E-200/HFC-E-202
Page 4 of 30
1. Installation and Operation
The Hastings HFM-E-200 mass Flow-meter and HFC-E-202 Flow-controller are designed to accurately
measure and control mass flow over the range of 10 sccm to 30 slm, without corrections or
compensations for gas pressure and temperature with an accuracy of better than ±1% FS. Hastings mass
flow instruments do not require any periodic maintenance under normal operating conditions with clean
gases. No damage will occur from the use of moderate overpressures (~500 psi/3.45MPa) or overflows.
Instruments are normally calibrated with the appropriate standard calibration gas (nitrogen) then a
correction factor is used to adjust the output for the intended gas
1.1. Features
LINEAR BY DESIGN. The HFM-E-200/HFC-E-202 series is inherently linear (no linearization circuitry is
employed). Should recalibration in the field be desired (a calibration standard is required), the
customer needs to simply set the zero and span points. There will be no appreciable linearity
change of the instrument when the flowing gas is changed.
MODULAR SENSOR. The HFM-E-200/HFC-E-202 series incorporates a removable/replaceable sensor
module. Field repairs to units can be achieved with a minimum of production line downtime.
LOW TEMPERATURE DRIFT. The temperature coefficient of span for the HFM-E-200/HFC-E-202
series is typically less than 0.05% of full scale/°C from 10-50°C. The temperature coefficient of zero
is typically less than 0.1 % of reading/°C from 10-50°C.
CURRENT LOOP. The 4-20 mA option gives the user the advantages of a current loop output to
minimize environmental noise pickup.