Datalogging
Manual Data Logging (999 sets)
1. Log readings one at a time to the internal memory by pressing the M MEMORY button.
“DATA M NO XXX” will appear on the LCD for one second to indicate the memory location.
2. Press R READ button to review logged readings. The display will show “DATA R NO XXX”.
3. Use the and buttons to scroll the logged readings.
4. Press R READ again to discontinue viewing the logged readings.
Continuous Data Logging
1. Press HOLD for 2 seconds and the display will show the INTV icon.
2. Use the or button to select the desired interval time (datalogging sample rate) from 1
second to 255 seconds.
3. Press SET to exit the interval time setting mode.
4. Press and hold M MEMORY for 2 seconds to enter the continuous (automatic) logging mode, the
display will show the M icon.
5. The M will flash each time a reading is stored.
6. Press M MEMORY again to exit the continuous datalogging mode.
7. Data stored using the continuous datalogging mode cannot be read directly on the tester’s
display, it must be downloaded to a PC using the supplied software.
Clearing the Datalogger Memory
When the internal memory is full, the Full icon will appear on the display and datalogging will stop.
1. Press
to switch OFF the tester.
2. Press and hold the MEMORY button, and while continuing to hold the MEMORY button, press
the
button. The display will show the CLr icon and all datalogged readings will be cleared from
memory.