Automatic(ATC)andManualTemperatureCompensation
ThePH210automaticallycompensates(ATC)forsolutiontemperaturedeviationswhenthe
Model850188temperatureprobe(insertedatbottomofmeter)isplacedinthe measurement
solution.Thetemperatureofasolutionisdisplayedonthemeter’slowerLCDline.
Tomanuallycompensateforsolutiontemperaturedeviationsaseparatethermometermustbe
placedinthesolutiontomonitorthesolution’stemperature;thenproceedasfollows:
1. PressandholdtheSETTINGbuttonfor5seconds.ThedisplaywillshownoATC.
2. PressENTERandnoATCwillflash(indicatingtheautomaticcompensationisswitchedOFF
andManualTemperatureCompensationisactive).Notethat
whennoATCisnotflashing
AutomaticTemperatureCompensationmodeisactive.
3. Usetheupanddownbuttonstomanuallyadjustthetemperaturevaluetomatchthe
solutiontemperature(asdeterminedbythesolutionthermometer).Oncethetemperature
isprogrammed,pressENTERtosavethedataandthenpressESC
toreturntothenormal
operatingmode.
4. Ifthesolutiontemperaturechanges,themanualtemperaturecompensationprocedure
mustberepeated.
TemperatureUnits(C/F)Selection
Tochangethetemperatureunits(
o
C/
o
F):
1. Poweronthedevice
2. PressandholdtheSettingbuttonfor5seconds,untiltheSettingmodeisaccessed
3. PresstheSettingbuttonagain;eitheranForaCwillbedisplayed
4. Usetheuparrowbuttontoselectthedesiredunitofmeasurefortemperaturesolution
5. PressENTERtosavethesetting
6. PressESCtoreturntothenormalmodeofoperation
AutomaticPowerOFFEnable/Disable
Toconservebatterypower,theinstrumentautomaticallyshutsoffafterapproximately10
minutesofinactivity.Todisableorenablethisfeature,followthestepsbelow:
1. Poweronthedevice
2. PressandholdtheSETTINGbuttonfor5secondsuntiltheSettingmodeisaccessed
3. UsetheSETTINGbutton
tosteptothescreenthatshowseitherOFFYESorOFFNO
4. UsetheuparrowbuttontoselectYES(AutoPowerOFFactive)orNO(AutoPowerOFFnot
active)