12 9
Fig. 5
3
20
22
Fig. 4
6
8
To be more mobile during oxygen inhalation at home (see Fig.
5), the 20 m connecting hose (22) can be screwed between
the pressure reducer (3) and the spiral hose (20).
time
available
xx
= 1143 min. = 19 hours, 03 min.
6.4 Filling level of the cylinder and running
time
The filling level of, for example, a 2 l oxygen cylinder is
calculated as follows:
cylinder volume x cylinder pressure = oxygen supply
2 l x 200 bar = 400 l
2 l x 100 bar = 200 l
The time available for oxygen inhalation is calculated as
follows:
oxygen supply x 4
=
cycle- switch- respiratory
delivery setting frequency per min.
Example:
With an oxygen supply of 400 l, the time available with a
respiratory frequency of 20 breaths per minute and a switch
setting of 2 is:
400 l x 4
0.035 l x 2 x 20
6.5 Alarm
● An audible alarm sounds after app. 45 sec. if OXYTRON
2
does no longer register any inhalation impulse. The red
alarm indicator starts flashing additionally.
6.6 Handling with Caddy
● Remove the shoulder strap
(18) from the bag (see
Fig. 11) and place it in the
OXYTRON
2 Mobile
carrying bag (1).
1
18
Fig. 11