A SERVICE OF

logo

7
VIIVII
VIIVII
VII.
General GuidelinesGeneral Guidelines
General GuidelinesGeneral Guidelines
General Guidelines
Calibration standards and sample solutions should
have the same ionic strength. ISA should be added to
both samples and standards immediately before tak-
ing measurements.
Calibration standards and sample solutions should be
stirred at the same rate using identical sized stir bars.
Stir thoroughly and continuously.
Calibration standards and sample solutions should be
at the same temperature. Thermally insulate solution
vessel from magnetic stirrer with cork or other insulat-
ing medium.
Wait until sensor value has stabilized before taking
reading (at least 5 minutes when going from more
dilute to more concentrated samples, longer when the
order is reversed).
Surface coating on the PTFE membrane will effect the
response. Inspect sensor before using. Wash off with a
jet of deionized water or mild detergent to remove
film.
Replace PTFE membrane if damage is evident or a
droplet of internal electrolyte is seen.
Rinse electrode with distilled or deionized water be-
tween samples and dab dry with lab wipe or other soft
disposable absorbent toweling.
Check calibration every 1-2 hours. Recalibrate if nec-
essary.
Position sensors at an angle of approximately 20° to
30° from verticle to reduce possibility of trapping gas
bubbles on on the membrane cap. Gas bubbles also
form from solution out-gassing due to temperature
change. Gently tap body of sensor to dislodge them.
Close container with plastic wrap or use HI 4000-71
gas sensor test vessel to prevent gas from leaving.
If electrode was left out in air for a prolonged period,
gently pulling cable will permit an exchange of fill
solution at the membrane surface. Re-Calibration is
required.