8
3. Holding the assembled electrode at the cable end,
shake the sensor to ensure internal fill solution that
may have separated during shipping is in contact
with inner membrane surface.
4. Prepare HI 5315 reference electrode by filling electro-
lyte reservoir with HI 7078 fill solution.
5. Place sensor and reference electrodes into electrode
holder and connect cable connectors to meter.
6. Soak the Nitrate electrodes membrane in a Nitrate
containing standard (0.001M) without ISA before
calibration.
HI 4113
The Hanna HI 4113 is shipped disassembled. The sensor is
shipped with two HI 4113-51 modules.
1. Unwrap Parafilm® seal found over ceramic junction
on inner stem and discard. This is only used for ship-
ping or long term storage.
2. Remove sensing cone from shipping vial. Do not touch
the sensing membrane with the H hole pattern on
it.
3. Screw the cone into the inner stem finger tight. Do not
over tighten.