Calibration standards and sample solutions should
have the same ionic strength. ISA should be added to
both samples and standards immediately before tak-
ing measurements.
Calibration standards and sample solutions should be
at the same temperature. Thermally insulate solution
vessel from magnetic stirrer with cork or other insulat-
ing medium.
Calibration standards and sample solutions should
be stirred at the same rate using identical sized stir
bars.
Surface coating or wetting the PTFE membrane will
effect the response. Inspect sensor before using. Re-
place PTFE membrane if damage is evident.
Rinse electrode with distilled or deionized water be-
tween samples and dab dry with lab wipe or other soft
disposable absorbent toweling.
Check calibration every 1-2 hours.
Position sensors at an angle of approximately 20° to
lessen bubble adherence from solution out-gassing
due to temperature change.
Close container with plastic wrap or use
HI 4000-71 gas sensor test vessel to prevent gas
from leaving.
Gently pulling cable will permit an exchange of fill
solution at membrane surface. Re-Calibration is re-
quired.
ions are capable of reducing the ammonia concentration.
If other volumes of sample/standard are used, add ISA at 2
parts per 100 parts standard/sample.