7. Cleaning and maintenance
Maintenance activity daily weekly monthly every
4 months
Inspect and, if necessary, √
clean wax bath strainers .
Inspect wax bath air vent and,
√
if necessary, clear.
Inspect and empty condensate bottle.
√
Clear inlet holes.
Examine activated carbon filter.
√
Replace carbon filter. √
R
emove fluid level sensor (float),
inspect and clean. √
Inspect electrical connectors √
at the rear panel of the instrument.
Inspect and, if necessary,
√
clean inside reagent bottles.
Check that the bottle caps and seals
√
(O-rings) of the reagent bottles are in
place and tight.
Check that the reagent bottles are fully
√
inserted in the connection manifolds.