NOTE How to use the preset function
1. To obtain an preset value, it is necessary to set up the reference gage in place
(the preset value is a compensation value determined from the measurement of
the reference gage). This preset setup takes about 1 second.
2. If the existing setup value is applied, it is not
necessary to carry out the preset. To force the
preset operation using the same data, move the
highlighted digit place with the key. This
makes the preset carried out, since the system
judges the data is changed.
For this operation press the
P.S V/
P.SET
key in the
ready state. With this single key operation preset
will be carried out.
3. So that the maximum displayable range is not exceeded during measurement, the
preset value must be set well within the maximum value shown in the table below.
If measured data exceed the maximum value, 9999999 will be displayed.
(noituloseR µ )m)mm(eulavmumixaM)E(noituloseR)E(eulavmumixaM
50.0/20.0/10.0 ± 99999.98
/200000./100000.
500000.
± 999999.8
5.0/2.0/1.0 ± 9999.998
/50000./20000./10000.
500./5000./2000./1000.
± 99999.98
001/01/5/2/1 ± 999.9998
4. Precautions prior to modifying the unit system
Note that if an preset value exceeds the above described maximum value when
the unit system is changed from E to metric, the preset function will automatically
be reset.
(Example: If the integer part of the maximum value is restricted to 2 digits,
converting from 4E to 101.6 mm will exceed the limit.)
5. To use the combined preset function together with the combined calibration
function for dual-unit measurement of a DF-type setup, set the direction to 0
(positive).
6. Preset setting is also possible with the following key entry:
SHIFT
,
P.S V/
P.SET
, (gage
size), and
ENT
.
7. If
P.S V/
P.SET
is entered in the ready state, zero-setting will be performed if no preset
has been made.
8. Preset/Zero-set can also be canceled with the following key entry:
SHIFT
,
P.S V/
P.SET
,
C
, and
ENT
.