74 AWG5000 Series Quick Start User Manual
7. Set Wait Trigger to the index number 1
using the Sequence Control Parame-
ters dialog box. This makes the
sequence wait a trigger before gener-
ating the first element.
8. Set Repeat Count to 5 at the index
number 2. This makes the sequence
repeat the second element five times.
9. Set Go To target to 1 at the index
number 3. This makes the sequence
jump to index number 1 after generating
the index number 3.
10. Click the Run button to confirm that the
sequence works properly. The anima-
tion window next to the Run button will
be activated when the instrument is in
the running state.
This screen image shows Table mode is
selected in the Sequence window.