![](https://pdfstore-manualsonline.prod.a.ki/pdfasset/1/9b/19baf8cb-9a11-4652-82e9-a8eaacde2517/19baf8cb-9a11-4652-82e9-a8eaacde2517-bg29.png)
Teledyne Advanced Pollution Instrumentation Model 633 Aethalometer
®
Black Carbon Monitor User Manual
07645A DCN6170 39
4.2InstrumentStatus
Theinstrument’scurrentstatusconditionisdisplayedindecimalformatonthefrontpanel
HOMEscreenandintheStatuscolumnofthedatadownload.Thestatusconditionrelatesto
varioussubcomponentsoftheinstrument,e.g.,Detector,Flow,LED,etc.Thedecimalnumber
representsasumofallofthestatusconditionsoccurringatanygiventime.(Multiplestatus
conditionsareinterpretedusingsubtractionofthelargestpossibleStatusConditionvalue
usingthetablebelow.
Table 1. Status Conditions and Descriptions.
Parameter
Status
Condition
Description
0 measuring
1 notmeasuring(duetoeithertapeadvance,fast
calibrationinprogress)
2 calibrating(LED,Flow,TapeSensors)
Detector
3 stopped
flowOK
4 flowlow/highbymorethan0.25LPM
8 Flowcheckstatushistory
Flow
12 FlowL/Handcheckstatushistory
LEDsOK
16 calibrating
32 calibrationerrorinoneormorechannels
(atleastonechannelOK)
LED
48 lederror(allchannelscalibrationerror,COMerror)
tapeadvanceOK
128 tapewarning(lessthan10spotsleft)
256 tapelastwarning(cardboxvisible,lessthan5spotsleft)
Tape
Advance
384 tapeerror(tapenotmoving,endoftape)
Notest
1024 Stabilitytest
2048 Cleanairtest
Tests
3072 Changetape
WhenStatusCondition3isencountered,theAethalometerstops.Inallotherstatuses,it
continuestooperatewithawarning,andthedataisflaggedaccordingly.Thestatusisrep‐
resentedbyonevalue,whichcanpointtooneparameteroracombinationofparameters.