Deltech Fitness 6 1 0 1 Marine Sanitation System User Manual


 
22
Mechanical Inspection
Press each key on the keyboard. Each key
should have a distinctive dome feeling. The
keys should not feel flat.
Attach the battery door. The battery door
should fit snugly in place when it is closed on
the pump.
Attach either a 50- or 100-ml Medication
Cassette reservoir or a CADD
®
administra-
tion set to the pump. Check for smooth
operation and a definite feel when the latch
pulls the reservoir or administration set firmly
against the bottom of the pump. The mark on
the latch should be aligned with the solid dot.
Lock the device by inserting a key into the
lock and turn counterclockwise until the
mark lines up with the solid dot.
Gently twist and pull on the cassette to make
sure it is firmly attached.
NOTE:
In the PCA delivery mode, the cassette must
be locked in order to start the pump.