14423-INS-LAB-RevD11 36
11APPENDIXB,EMCINFORMATION
GUIDANCEANDMANUFACTURER’S
DECLARATION
ElectromagneticEmissions—forallEquipment
andSystems
Guidanceandmanufacturer’sdeclaration—
electromagneticemission
The Grafco Ultrasound Pocket Doppler is intended for use in the
electromagnetic environment specied below. Do not use the system in
environments which do not comply with the specications listed below.
Emission
test
Compliance Electromagneticenvironment–
guidance
RF emission
CISPR 11
Group 1 The Grafco Ultrasound Pocket Doppler
uses RF energy only for its internal function.
Therefore, its RF emissions are very low
and are not likely to cause any interference
in nearby electronic equipment.
RF emission
CISPR 11
Class B The Grafco Ultrasound Pocket Doppler
is suitable for use in all establishments,
including domestic establishments and
those directly connected to the public
low-voltage power supply network that
supplies buildings used for domestic
purposes.
ElectromagneticImmunity—forallEquipment
andSystems
Guidanceandmanufacture’sdeclaration—electromagnetic
immunity
The Grafco Ultrasound Pocket Doppler is intended for use in the
electromagnetic environment specied below. Do not use the system in
environments which do not comply with the specications listed below.
Immunity
test
IEC60601
testlevel
Compliance
level
Electromagnetic
environment—guidance
Electrostatic
discharge
(ESD) IEC
61000-4-2
±6 kV
contact ±8
kV air
±6 kV
contact ±8
kV air
Floors should be wood,
concrete or ceramic tile. If
oor is covered with synthetic
material, the relative humidity
should be at least 30% RH.