Teledyne 200EU-NOy Option Manual Addendum Marine Battery User Manual


 
Model 200EU-NOy Option Manual Addendum
SECONDARY MENU (The fixed characteristics of the system).
PRESS DISPLAY ACTION
SEL HOLD TILL p-n1
SET TO 0 – REVERSE ACTION
SEL HOLD TILL p-n2
SET TO 2 = TYPE J THERMOCOUPLE
SEL
p-dF DIGITAL FILTER
SET TO 5
SEL
P-SL LOWER LIMIT
DEFAULT = 32
SEL
P-SU UPPER LIMIT
SET TO 350 (350ºC)
SEL
P-Ab ALARM
NONE
SEL
P-An HYTERESIS
NONE
SEL
P-dP PROG DECIMAL
SET TO 0
SEL
P-48 COLD JUNCTION
SET TO 2 (TYPE J)
SEL
r5 RAMP TO SET POINT
LEAVE AT 0
SEL
PVOF PROCESS OFFSET
LEAVE AT 0
SEL
SVOF SET POINT OFFS
LEAVE AT 0
SEL
P-F
SET TO 0 (0 =
o
C, 1 =
o
F)
SEL
P-n1
SEL
HOLD TILL P
SEL
SCROLL TO A7
SET TO 1 (AUTOTUNE )
SV SET VALUE
SHOULD BE 315°C
05386 Rev. A 31