Conductivity(TDS&Salt)Calibration
Conductivityaccuracyverificationshouldbeperformedonaperiodicbasis.Oncepermonthis
therecommendedcyclefornormaluse.Ifcalibrationisrequired,aconductivitystandardizing
solutionmustbeobtained.Themetercanbecalibratedinanyorallofthethreeranges.
Standardizingsolutionsof84µS/cm,1413µS/cmor12.88mS/cm
(12,880µS/cm)areusedforthe
automaticcalibrationrecognitionprocedure.Noothercalibrationvaluesarepermitted.
Calibrationisalwaysdoneinconductivitymode.SincesalinityandTDSvaluesarecalculated
fromconductivityvalues,thisprocedurealsocalibratestheSalinityand TDSranges.
1. Fillasamplecupwiththestandardizingsolution.
2. TurnthemeterONandinserttheelectrodeintothesolution.Tapormovetheelectrodein
thesampletodislodgeanyairbubbles.
3. PressandholdtheMODE/CALbutton(approximately2seconds)untilthemaindisplay
startsflashing.
4. Themeterwillautomaticallyrecognizeandcalibratetothe
standardizingsolution.Atthe
endofthecalibration,thedisplaywillbrieflyindicate“SA”,“End”andthenreturntothe
measurementmode.
Note:The“SA”willnotappearifthecalibrationfails.
5. The“rangecalibrated”symbolwillappearinthedisplayforeachrangethatiscalibrated
duringthat
poweroncycle.
Lowrange,84µS/cm
Mediumrange,1413µS/cm
Highrange,12.88mS/cm(12,880µS/cm)
Note:Themeterallowsfora1,2or3pointcalibration.Ifcalibrationisdoneformorethanone
point,thelowestvaluestandardshouldbedonefirsttoobtainthebestaccuracy.
ORPCalibration
TheORPelectrodedoesnotrequirecalibration.
Temperature(pHandConductivity)OffsetCalibration
Thisprocedureallowsforerrorcorrectionoftheexternaltemperatureprobe(pH)orthe
conductivityprobe’sbuiltintemperaturesensor.
1. SwitchtopHorConductivity(Salt/TDS)mode.
2. Placethetemperatureprobeorconductivitycellinthesampleandallowthetemperature
readingtostabilize.
3. PressandholdtheUNIT/SET
buttonuntiltheºCorºFiconbeginsflashing.
4. Adjusttheorbuttonstosetthedisplaytoindicatetheknowntemperatureofthe
sample.
5. MomentarilypresstheUNIT/SETbuttontosavethechangeandreturntothemeasurement
mode.