To be more mobile during oxygen inhalation at home (see Fig.
5), the 20 m connecting hose (22) can be screwed between
the pressure reducer (3) and the spiral hose (20).
time
available
xx
= 1143 min. = 19 hours, 03 min.
6.4 Filling level of the cylinder and running
time
The filling level of, for example, a 2 l oxygen cylinder is
calculated as follows:
cylinder volume x cylinder pressure = oxygen supply
2 l x 200 bar = 400 l
2 l x 100 bar = 200 l
The time available for oxygen inhalation is calculated as
follows:
oxygen supply x 4
=
cycle- switch- respiratory
delivery setting frequency per min.
Example:
With an oxygen supply of 400 l, the time available with a
respiratory frequency of 20 breaths per minute and a switch
setting of 2 is:
400 l x 4
0.035 l x 2 x 20
6.5 Alarm
An audible alarm sounds after app. 45 sec. if OXYTRON
2
does no longer register any inhalation impulse. The red
alarm indicator starts flashing additionally.
6.6 Handling with Caddy
Remove the shoulder strap
(18) from the bag (see
Fig. 11) and place it in the
OXYTRON
2 Mobile
carrying bag (1).
1
18
Fig. 11