![](https://pdfstore-manualsonline.prod.a.ki/pdfasset/6/30/630b184d-2030-4b0c-b8f6-cd36e28d8048/630b184d-2030-4b0c-b8f6-cd36e28d8048-bg45.png)
Usin
your Pro
rammable Power Supply 5 - 17
Local control (in CV or CC mode):
The volta
e V
set
can be set usin
the key and the numeric input keys.
Example:
Press the V key to start the volta
e settin
.
Press the keys 3, . (dot), 4 and ENTER to enter the settin
of 3.4V.
The current I
set
can be set usin
the key and the numeric input keys.
Example:
Press the I key to start the current settin
.
Press the keys 0, dot (.), 2, 3 and ENTER to enter the settin
of 0.23A.
The overvolta
e protection level OVP
set
can be set usin
the key and
the numeric input keys.
Example:
Press the OVP key to initiate the overvolta
e settin
.
Press the keys 7 and ENTER to enter the overvolta
e settin
of 7V.
The overcurrent protection can be enabled/disabled usin
the key.
Example:
Press the OCP EN/OCP DIS key.
IF OCP was enabled THEN OCP is disabled
ELSE OCP is enabled.
IF OCP has been enabled, the annunciator above the text will be displayed.
An overcurrent or overvolta
e trip protection can be reset by pressin
the
key. The display of OVERCURRENT or OVERVOLTAGE will also
disappear.
In the CV mode, the annunciator above the text CV will be displayed.
In the CC mode, the annunciator above the text CC will be displayed.
OCP
EN