• keep power cord visible and accessible at all times. e plug on the power cord serves as the
disconnect device. e wall socket outlet shall be as close as practical and shall be easily
accessible.
• use only the power cord specied for this product and certied for the country of use.
• do not allow the power cord to get wet.
• do not use the warming unit when it appears the warming unit, power cord or
any component is damaged. Contact 3M Patient Warming Technical Support at
1-800-733-7775.
• this equipment must only be connected to a supply mains with protective earth.
12. To reduce the risks associated with exposure to biohazards always perform the
decontamination procedure prior to returning the warmer for service and prior to disposal.
13. Do not retain the patient using a warming blanket alone, as injury may result. Use a draw,
sheet safety strap, or other means to retain the patient.
14. Do not modify this equipment without authorization of the manufacturer.
15. To ground the Bair Hugger warming unit, only connect to receptacles marked “Hospital
Only,” “Hospital Grade,” or a reliable grounded outlet.
CAUTIONS
1. Except for specific blanket models, Bair Hugger blankets are not sterile and are all intended
for single patient use ONLY. Placing a sheet between the Bair Hugger blanket and the patient
does not prevent contamination of this product.
2. Monitor the temperature and cutaneous response of patients who are incapable of reacting,
communicating and/or who are without a sense of feeling every 10-20 minutes or according
to institutional protocol. Monitor the patient’s vital signs regularly. Adjust air temperature
or discontinue therapy when the therapeutic goal is reached or if vital sign instability occurs.
Notify physician of vital sign instability immediately.
3. Do not leave pediatric patients unattended during therapy.
4. Do not initiate temperature management therapy unless the temperature management unit
is free from mechanical damage and is safely placed on a hard surface or securely mounted.
Otherwise, injury may result.
5. To prevent tipping, clamp the Model 775 temperature management unit to an IV pole at a
height that provides stability. We recommend clamping the unit no higher than 44” (112 cm)
on an IV pole with a minimum 28” (71 cm) diameter wheelbase. Failure to do so may result in
IV pole tipping, catheter site trauma, and patient injury.
6. Electrical shock hazard. Do not disassemble the temperature management unit unless you
are a qualified service technician. There are electrically live parts within the unit when it is
connected to a power source, even when the unit is in Standby mode.
7. To reduce the risks associated with environmental contamination follow applicable
regulations when disposing of this device or any of its electronic components.