Safety System
Independent bulb and capillary
Dual fused input lines
Over-heat: flashing red light with audible alarm;
heater shuts down
IEC 60601-1; IEC/EN 60601-1-2; UL 60601-1;
CAN/CSA-C22.2, No.601.1; EN 55011; EN 80601-
2-35.
MEDICAL — GENERAL MEDICAL
EQUIPMENT AS TO ELECTRICAL SHOCK,
FIRE AND MECHANICAL HAZARDS ONLY IN
ACCORDANCE WITH UL 60601-1; CAN/CSA-
C22�2, No�601�1; ANSI/AAMI ES60601-1:2005
CSA-C22�2 No� 60601-1:08; EN 80601-2-35;
Control No�4HZ8
Classified under IEC 60601-1 Guidelines (and other
national versions of the Guidelines) as Class I, Type
BF, Ordinary equipment, Continuous operation.
Not suitable for use in the presence of flammable
anesthetic mixtures with air or with oxygen or nitrous
oxide. Classified by Underwriters Laboratories Inc.
with respect to electric shock, fire and mechanical
hazards only, in accordance with UL 60601-1, EN
80601-2-35 and in accordance with Canadian/CSA
C22.2, No. 601.1. Classified under the Medical
Device Directive as a Class IIb device.
Electrical Characteristics
Operating speed: 3150 rpm
Airflow: 28-30 cfm
Peak: 1000W
Average: 450W
<100µA
850W Resistive
15-foot, SJT, 3 cond., 10A
4.6 meter, HAR, 3 cond., 10A
110-120 VAC, 60 Hz, 9.5 Amperes, or
220-240 VAC, 50 Hz, 4.5 Amperes, or
100 VAC, 50/60 Hz, 10 Amperes
10A, 200mA (110 - 120 VAC Units)
6.3A, 100mA (220 - 240 VAC Units)
15A, 160mA (100 VAC Units)
Over-heat test can be performed by the biomedical
group.
14 English Bair Hugger Model 505 Temperature Management Unit Operator’s Manual
Check the 3M
TM
Bair Hugger
TM
therapy website to ensure you have the most recent version of this document. www.bairhugger.com reorder #202431A