Campbell Hausfeld CS115 Marine Instruments User Manual


 
CS115 Barometric Pressure Sensor
8
8: Sample (P70)
1: 1 Reps
2: 1* Loc [ P_mb ]
*Table 2 Program
02: 0.0000 Execution Interval (seconds)
*Table 3 Subroutines
End Program
-Input Locations-
1 P_mb
* Proper entries will vary with program and datalogger channel, and input location assignments.
5.1.2 Period Averaging Measurement Example for CR10(X) in a Slow
Executing Program
EXAMPLE 2. Sample Program for CR10(X)
;{CR10X}
;
*Table 1 Program
01: 10 Execution Interval (seconds)
;Turn on CS115
;
1: Do (P86)
1: 48* Set Port 8* High
;Give 2 second delay for CS115 to settle
;
2: Excitation with Delay (P22)
1: 1 Ex Channel
2: 0 Delay W/Ex (0.01 sec units)
3: 200 Delay After Ex (0.01 sec units)
4: 0 mV Excitation
3: Period Average (SE) (P27)
1: 1 Reps
2: 14 200 kHz Max Freq @ 2 V Peak to Peak, Freq Output
3: 6* SE Channel
4: 10 No. of Cycles
5: 5 Timeout (units = 0.01 seconds)
6: 1* Loc [ P_mb ]
7: 1000 Mult
8: 0.0 Offset