Emerson 6081-P Marine Instruments User Manual


 
6
6081-P Instruction Sheet
October 2013 LIQ_MAN_ABR_6081-P
EEPROM Warning
Sense Line Open Warning
Factory Cal Warning
Keyboard Warning
Once a fault or warning is detected, the display will show a message describing the problem.
Sensor Temperature Range: -10 to 150°C (PT100 and PT1000)
Display: 2-line, 16 character display supports display of pH and mV units. Display shows temperature.
Approvals:
RFI/EMI:
EN-61326
EN 301 489-1 V1.2 2002
EN 301 489-17: V1.4.1 2002
EN 60950-1: 2001
EN 300 328 V 1.6.1 (2004-11)
Hazardous Location Approvals
Intrinsic Safety:
Special Conditions of Use:
1. The 6081 enclosure is made of aluminum alloy and
given a protective polyurethane paint nish; however,
care should be taken to protect it from impact or
abrasion if located in zone 0. This applies to any
Intrinsically Safe installation, whether the Class, Division
or Zone system is used.
2. Potential Electrostatic Hazard – The plastic antenna may
present a potential electrostatic ignition hazard and
must not be rubbed or cleaned with a dry cloth.
Intrinsically Safe,
Class I, Division 1
Groups ABCD/T4
Ta = -20 °C to +65 °C – 1400322;
IP66
Class 1 Zone 0, AEx ia IIC T4
Ta = -20 °C to +65 °C – 1400322;
IP66
Non-Incendive:
Nonincendive, Class I,
Division 2, Groups ABCD/T4
Ta = -20˚C to +65 °C
Complies with FM standards:
FM3600:1998
FM3610: 2010, ISA60079-0:2009, ISA60079-11:2009.
FM3611: 2004
FM3810: 2005
ANSI/IEC 60529:2004
Class I, Division 1, Groups A/B/C & D
Class II, Division 1, Groups E/F& G
Class III
T4 Tamb : -20 to +65 °C
Type 4x, IP66
CE 1180 II1G
Baseefa 10 ATEX 0149X
Ex ia IIC T4 Ga (-20 °C ≤ Ta ≤ +65 °C
(source CE cert-6081 Baseefa)
Complies with Standards: EN
60079-0:2009, EN 60079-11:2007
Class I, Division 2,
Groups A, B, D
Dust Ignition Proof
Class II, Division 2,
Groups F & G
T4 Tamb : -20 to +65 °C
Environmental: The operating atmosphere of the
transmitter must be consistent with the appropriate