SECTION II
EQUIPMENT REQUIREMENTS
4. Equipment Required. Table 2 identifies the specific equipment to be used in this
calibration procedure. This equipment is issued with Secondary Transfer Calibration
Standards Set AN/GSM-287 or AN/GSM-705. Alternate items may be used by the
calibrating activity. The items selected must be verified to perform satisfactorily prior
to use and must bear evidence of current calibration. The equipment must meet or
exceed the minimum use specifications listed in table 2. The accuracies listed in table 2
provide a four-to-one ratio between the standard and TI.
5. Accessories Required. The accessories required for this calibration are common
usage accessories, issued as indicated in paragraph 4 above, and are not listed in this
calibration procedure. The following peculiar accessories are also required for this
calibration: Adapter, SMC jack to BNC plug, Hewlett-Packard Model 1250-0831;
Extender Boards, 10-terminal circuit board extenders (05345-60201 and 05345-60202);
and Extender Cable, Hewlett-Packard Model 05345-60205.
Table 2. Minimum Specifications of Equipment Required
Common name
Minimum use specifications
Manufacturer and model
(part number)
AUTOTRANSFORMER Range: 105 to 125 V ac
Accuracy: ±1%
General Radio, Type
W10MT3AS3 (7910809) or
Ridge, Model 9020A (9020A)
or Ridge, Model 9020F (9020F)
FUNCTION GENERATOR Range: 10 Hz to 30 MHz
Amplitude: 0 to 40 mV rms
Accuracy: ±3% (±1.5 dB)
(SG-1288/G)
FREQUENCY DIFFERENCE
METER
Range: 10 MHz
Resolution: 1 part in 10
-10
per day
3.7 to 12.4 GHz, Hewlett-
Packard, Model 537A
(7910718-2)
MULTIMETER Range: -2 to +2 Vdc
Accuracy: ±0.000125%
Range: 0 to ± 15 V dc
Accuracy: ±0.03%
Hewlett-Packard, Model
3458A (3458A)
OSCILLOSCOPE Bandwidth: 100 MHz
Sensitivity: 50 mV/div
Sweep rate: 10 ns/div
Accuracy: ±3%
Tektronix, Type 2465BOPT46
(2465BOPT46)
SIGNAL GENERATOR Range: 30 to 500 MHz
Amplitude: 0 to 40 mV rms
Accuracy: ±3% (±1.5 dB)
(SG-1207/U)
TIME/FREQUENCY
WORKSTATION
Frequency: 10 MHz
Accuracy: 1.25 parts in 10
-10
per day
Autek Systems Corp. Model
620 (MIS-38946)