Calibration Factor (CAL)
The Calibration factor (CAL) provides a means to improve the accuracy of the results display by calibrating
against the output of a known frequency generator. The field strength value measured is multiplied by the
value of CAL that has been entered and the resulting value is displayed. The CAL setting range is from 0.20
to 5.00. The measurement accuracy will be sufficient for most applications even if the frequency response of
the sensor calibration factor is ignored and the CAL is set to 1.00 in at all points.
Setting the calibration factor (CAL)
Hold-on
Power off
+
When the meter is normally turned on, the calibration factor set value will display for 2 seconds.
1. Press
key to turn off the meter.
2. Press and hold on the XYZ/CAL key, then press
key to turn on the meter to enter the
calibration factor setting mode, the “ CAL SEt ” marks is displayed.
3. Press or key to increase or decrease the value.
4. Press the SET key to store the new setting value and exit.
E-Field Typical Calibration Data:
Frequency CAL Frequency CAL
50MHz 2.13 900 MHz 1.40
100MHz 1.37 1.8GHz 2.06
200 MHz 1.19 2.4GHz 0.66
300 MHz 0.69 3.5GHz 1.05
433 MHz 0.78 5.4GHz 2.20
500 MHz 1.38 8.0GHz 3.16
600 MHz 2.12
700 MHz 1.66
800 MHz 1.40