26) A review screen will appear with three options:
ABORT - do NOT save the data point, and return to Immediate
Mode to re-test the same data point. If a FAIL reading appears, it
is recommended to ABORT, clean all connections, and then re-
test.
RENAME - change the descriptive fiber name (same as step 27).
SAVE - save the data point, and advance to the next data point in Immediate Mode. Notice
that the fiber number automatically increments to the next number.
27) Once the data has been stored for the current fiber, move the units to the next fiber, and repeat steps
24 through 26 until all fibers have been tested.
Once all data is stored in the FiberMeter, they can be downloaded to a PC which has Extech Reporter
installed. Please consult the Extech Reporter unit for more information.
NOTE: the memory in the FiberMeter allows for permanent storage of data, including reference and power
readings. Data will remain in the meter, even when the unit is powered off, until it is removed by the user.