Emerson 1066 Marine Instruments User Manual


 
QUICK START GUIDE – 1066 Liquid Analytical Transmitter
1. For mechanical installation instructions, see page 8 for panel mounting and page 9 for pipe or wall mounting.
2. Wire the sensor to the main circuit board. See page 10 for wiring instructions. Refer to the sensor instruction
sheet for additional details. Make loop power connections.
3. Once connections are secured and verified, apply DC loop power to the transmitter.
4. When the transmitter is powered up for the first time, Quick Start screens appear. Quick Start operating tips are
as follows:
a. A highlighted field shows the position of the cursor.
b. To move the cursor left or right, use the keys to the left or right of the ENTER key. To scroll up or down or to
increase or decrease the value of a digit use the keys above and below the ENTER key. Use the left or right keys
to move the decimal point.
c. Press ENTER to store a setting. Press EXIT to leave without storing changes. Pressing EXIT during Quick Start
returns the display to the initial start-up screen (select language).
5. Choose the desired language and press ENTER.
6. Choose measurement and press ENTER.
a. For pH or ORP, choose preamplifier location. Select Analyzer to use the integral preamplifier in the transmitter;
select Sensor/J-Box if your sensor is SMART or has an integral preamplifier or if you are using a remote pream-
plifier located in a junction box.
5. If applicable, choose units of measurement.
6. For contacting and toroidal conductivity, choose the sensors type and enter the numeric cell constant using the keys.
7. Choose temperature units: °C or °F.
8. After the last step, the main display appears. The outputs are assigned to default values.
9. To change output settings, to scale the 4-20mA current outputs, to change measurement-related settings from
the default values, and to enable pH diagnostics, press MENU. Select Program and follow the prompts. Refer to
the appropriate menu.
10. To return the transmitter to the factory default settings, choose Program under the main menu, and then scroll to
Reset.
11. Please call the Rosemount Analytical Customer Support Center at 1-800-854-8257 if you need further support.
2
1066 Instruction Sheet
July 2013 LIQ_MAN_ABR_1066-CT-HT-FF-FI