Emerson 1066 Marine Instruments User Manual


 
6
1066 Instruction Sheet
July 2013 LIQ_MAN_ABR_1066-CT-HT-FF-FI
Specifications CONTINUED
TOROIDAL CONDUCTIVITY
Performance Specifications
Measurement Range: see table below
Input filter: time constant 1 - 999 sec, default 2 sec.
Response time: 3 seconds to 95% of final reading
Recommended Sensors for Conductivity
All Rosemount Analytical submersion/immersion and
flow-through toroidal sensors.
Temperature range -25 to 210°C (-13 to 410ºF)
Temperature Accuracy,
Pt-100, -25 to 50 °C
± 0.5°C
Temperature Accuracy,
Pt-100,. 50 to 210°C
± 1°C
Temperature specifications:
PERFORMANCE SPECIFICATIONS
Recommended Range - Toroidal Conductivity
Model 226: ±1% of reading ±5mS/cm in recommended range
Models 225 & 228: ±1% of reading ±15
mS/cm in recommended range
Models 222, 242: ±4% of reading ±5mS/cm in recommended range
Models 225, 226 & 228: ±5% of reading outside high recommended range
LOOP PERFORMANCE (Following Calibration)
Model 1mS/cm 10mS/cm 100mS/cm 1000mS/cm 10mS/cm 100mS/cm 1000mS/cm 2000mS/cm
5mS/cm to 500mS/cm
15mS/cm to 1500mS/cm
500mS/cm to 2000mS/cm
500mS/cm to 2000mS/cm
100mS/cm to 2000mS/cm
1500mS/cm to 2000mS/cm
226
242
222
(1in & 2in)
225 & 228