Emerson 1066 Marine Instruments User Manual


 
5
Instruction Sheet 1066
LIQ_MAN_ABR_1066-CT-HT-FF-FI July 2013
Specifications CONTINUED
CONTACTING CONDUCTIVITY
Performance Specifications
Measurement Range: see table below
Input filter: time constant 1 - 999 sec, default 2 sec.
Response time: 3 seconds to 95% of final reading using
the default input filter
Recommended Sensors for Conductivity
All Rosemount Analytical ENDURANCE Model 400 series
conductivity sensors (Pt 1000 RTD) and Model 410VP 4-
electrode sensor.
Temperature range 0 to 200°C
Temperature Accuracy,
Pt-1000, 0-50°C
± 0.1°C
Temperature Accuracy,
Pt-1000, Temp. > 50°C
± 0.5°C
Temperature specifications:
Linearity for Standard Cable
50 ft (15 m)
±0.6% of reading in recommended range
±2% of reading outside high recommended range
±5% of reading outside low recommended range
±4% of reading in recommended range
PERFORMANCE SPECIFICATIONS
Recommended Range – Contacting Conductivity
Cell 0.01S/cm 0.1mS/cm 1.0mS/cm 10mS/cm 100mS/cm 1000mS/cm 10mS/cm 100mS/cm 1000mS/cm
Constant
0.01
0.1
1.0
4-electrode
0.01mS/cm to 200mS/cm
0.1mS/cm to 2000mS/cm
1 mS/cm to 20mS/cm
2mS/cm to 1400mS/cm
200mS/cm to 2000mS/cm
2000mS/cm to 20mS/cm
20mS/cm to 200mS/cm