11
ON again and carefully turn the control knob until you feel a tingling or
slight twitch under or around the electrodes.) Always start with the low-
est setting and increase the intensity slowly.
If the stimulation levels are uncomfortable or become uncomfortable,
reduce the stimulation intensity to a comfortable level; or cease stimula-
tion and contact your physician.
6. Setting the Patient Compliance Counter:
a) To turn on the Patient Compliance Counter: While the unit is ON,
hold down the UP key and press the MODE key simultaneously.
b) To reset the Patient Compliance Counter: While the unit is ON, press
the UP key and the MODE key simultaneously (this will take you into
the Compliance Counter), then push the DOWN key and press the
MODE key simultaneously.
c) Press the UP and MODE keys simultaneously to return to the treat-
ment status.
BATTERY INFORMATION
A 9-volt battery is provided with your unit. When the LCD (Liquid Crystal
Display) low battery mark illuminates, the battery has become too weak to
power the unit, and the existing battery should be replaced with a new bat-
tery. At this point, the unit will turn OFF until a new battery is inserted.
Replacing the Battery
When the LCD low battery mark illuminates, and the unit does not remain illu-
minated once turned on, the battery should be replaced.
1. Turn unit OFF.
2. Remove the front panel cover by pressing on the top of the panel and
pressing down in order to slide the panel down. Continue sliding the
panel downwards until the panel is completely removed from the unit.
This will reveal the battery compartment.
3. Remove the discharged battery from the device.
4. Place new battery in the compartment. Note: Be sure the proper polarity
(+ and –) markings match the markings in the device.