14
LCD (Liquid Crystal Display)
The LCD displays timer (10, 20, 30, 40, 50, 60 minutes and C continuous),
function modes (S, C, A, Cycle ON / Ramp / OFF time), pulse width, pulse
rate, low battery, and patient compliance counter.
1. To check the LCD function, turn device ON and all the parameters will be
displayed for 1~2 seconds.
2. After 1~2 seconds, the LCD will go to previous operation parameter.
Changing the LCD Parameters
To set the treatment parameters on the EMS device, the unit must be
in the UNLOCK position. If the device is in the LOCK position it, can be
unlocked by pressing the UP and DOWN keys simultaneously for 2~3
seconds.
To switch between the different treatment parameters, press the MODE
key.
Once in the desired mode, press the UP, DOWN, or SET key until the
desired treatment parameter is obtained.
Once all of the desired treatment settings are displayed, the unit may
be locked by pressing the UP and DOWN keys simultaneously for 2-3
seconds while in the timer function.