Graham Field GF-TX5EMS SONAR User Manual


 
8
EXPLANATION OF KEY / KNOB CONTROL FUNCTIONS
DOWN Key This key decreases the setting
of treatment parameters. *
This key regulates the number of pulse width,
pulse rate, and cycle on/ramp/off time.
*Decrease by pressing the DOWN key: The width can be adjusted by 10μ/step.
The rate can be adjusted in 1Hz~20Hz by 1Hz/step, 20Hz~150Hz by 5Hz/step.
The cycle on/ramp/off time can be adjusted by 1 second/step.
MODE Key
This key selects Timer Mode,
Stimulation Mode,
Pulse Width Mode,
Pulse Rate Mode, or
Cycle On/Ramp/Off Time
This key changes the treatment parameter.
Each time the MODE key is pressed, the next
parameter will display. The selected treatment
parameter in the current mode will ash.
SET Key
This key switches between
the different settings within
Timer Mode and
Stimulation Mode.
Each time the SET key is pressed, the
parameter will change to the next setting
within the parameter. The selected setting will
ash. When the desired setting is ashing,
press the MODE key to conrm the choice.
UP Key This key increases the setting
of treatment parameters. *
This key regulates the number of pulse width,
pulse rate, and cycle on/ramp/off time.
*Increase by pressing the UP key: The width can be adjusted by 10μs/step.
The rate can be adjusted in 1Hz~20Hz by 1Hz/step, 20Hz~150Hz by 5Hz/step.
The cycle on/ramp/off time can be adjusted by 1 second/step.
Ch1 / Ch2
Knobs
Channel 1 and Channel 2
Intensity Control Knobs
These knobs control the strength of the
stimulation and also function as ON / OFF
controls.