Teledyne 2002 Marine Instruments User Manual


 
149-072011 Model 2002 Vacuum Gauge Page 32 of 38
The 0-10V output option board, provides voltage outputs proportional to the HPM-2002 pressure reading.
The first channel corresponds to the higher-pressure range (0-1024) Torr). The second channel
corrresponds to the lower pressure range (0-1000mTorr). The equation below gives the output voltage for a
given pressure.
Where:
V (channel 1) is the voltage between pins 1 and 3,
V (channel 2) is the voltage between pins 8 and 7.
P is the indicated pressure.
()
(
)
100
1
TorrP
channelV =
()
(
)
100
2
mTorrP
channelV =