A SERVICE OF

logo

<1>CUSTOM WAVELENGTH - the FiberMeter has the capability of setting a
custom wavelength. The custom wavelength temporarily replaces 980nm, and
requires a singlemode light source tuned to -10dBm for calibration.
Enter the 3- or 4-digit custom wavelength between 700nm and 1700nm in the
entry field provided, then press to continue. An example of a custom
wavelength would be 1490nm.
Connect a -10 dBm singlemode light source of the appropriate wavelength to the
FiberMeter using a singlemode patch cord.
Press to confirm calibration. The meter will then return to the METER
CONFIG MENU.
<2>DEFAULT WAVELENGTH - this WAVELENGTH OPTION resets a
previously set custom wavelength to the default wavelength of 980nm.
Press to confirm the default wavelength. The meter will then return to the
METER CONFIG MENU.
NOTE: if the wavelength is already set to 980nm, pressing the button has no
effect.
Entering Custom Wavelength
1-6
Setting Custom Wavelength
Resetting Custom Wavelength to Default
FEATURES & FUNCTIONS
UNIT 1
Set Custom
Wavelength
(
[ ]
WAVELENGTH(nm):ENTER
_____________________
<--- --->SHIFT
700 to 1700nm )
980nm
CONNECT
YES
SOURCE
1490nm
_____________________
WILL
REPLACE
1490nm
-10dBm
TO CALIBRATE
NO
Verify Custom
Wavelength
Restore Custom
Wavelength
RESTORE
YES
_____________________
980nm
NO
DONE
F1F1
F1F1
2
DEF