A SERVICE OF

logo

4-7
Extech Reporter
UNIT 4
Light Source Reference Power
Definition: the amount of absolute optical power being received directly from the light source by the
power meter during the setting of the optical reference. Light Source Reference Power is used as a
reference, or starting point, on which a standards-based optical reference is set.
Units: dBm
Calculation: none; Light Source Reference Power is measured via a single patch cord of the appropriate
type.
NOTE: testing standards require multimode reference
cables to be wrapped around a device called a mandrel,
which is used to filter high-order modes and achieve EMD,
or Equilibrium Mode Distribution (shown here).
Singlemode test jumpers do NOT require this mandrel.
Light
Source
Reference
Power (dBm)
INVALID
OPERATING
MARGIN **
Minimum
Required
Power (dBm)
Measured
Power (dBm)
PASS
OPERATING MARGIN
or
SYSTEM OVERHEAD
Optical Loss
FAIL
Total Allowable System Loss
Interpreting Extech Reporter Screens and Reports, cont.
CHARGER
850nm 1300nm
MULTI MODE
LIGHT SOURCE
DualDual
LEDLED