Emerson Process Management 402 Marine Instruments User Manual


 
402 and 402VP Instruction Manual
January 2014 LIQ_MAN_ABR_402_402VP
Rev. L
Credit Cards for U.S. Purchases Only.
8
Emerson Process Management
2400 Barranca Parkway
Irvine, CA 92606 USA
Tel: (949) 757-8500
Fax: (949) 474-7250
rosemountanalytical.com
© Rosemount Analytical Inc. 2013
©2014 Rosemount Analytical, Inc. All rights reserved.
The Emerson logo is a trademark and service mark of Emerson Electric Co. Brand name is a mark of one
of the Emerson Process Management family of companies. All other marks are the property of their
respective owners.
The contents of this publication are presented for information purposes only, and while effort has been
made to ensure their accuracy, they are not to be construed as warranties or guarantees, express or
implied, regarding the products or services described herein or their use or applicability. All sales are
governed by our terms and conditions, which are available on request. We reserve the right to modify
or improve the designs or specications of our products at any time without notice.
facebook.com/EmersonRosemountAnalytical
AnalyticExpert.com
twitter.com/RAIhome
youtube.com/user/RosemountAnalytical
Figure 15. Checking Temperature Element
Disconnect leads and measure resistances shown. The mea-
sured resistance should be close to the value in the table.
Figure 16. Checking Continuity and Leakage
Disconnect electrode leads and measure resistance and
continuity as shown. Sensor must be dry when checking
resistance between electrode leads.