Teledyne SERIES 4030 Scuba Diving Equipment User Manual


 
Installation
Teledyne Analytical Instruments 20
Both the first and third set of outputs as well as the second and
fourth set of outputs may be set to any of the following functions:
- THC -- Total hydrocarbon as measured in the first phase
of detection. The output is the percentage of the selected
range.
- SMP -- An alternate measurement that occurs during the
second phase of detection. The output represents the
percentage of the selected range.
- WAVE – Analogous to the continuous output of the FID.
- RID – Indicates the selected range. Low Range is at 1/5
of maximum output. Mid Range is at 1/2 of maximum
output. High Range is 4/5 of maximum output.
Table 3-1: Analog Output Connections
Pin Function
3 Channel 2, + 4-20 mA, floating
4 Channel 2, – 4-20 mA, floating
5 Channel 1, + 4-20 mA, floating
6 Channel 1, – 4-20 mA, floating
8 Channel 2, 0-1 VDC
23 Channel 2, 0-1 VDC
24 Channel 1, 0-1 VDC
7 Channel 1, negative ground