16 Troubleshooting
English 202052A
Errors
Error Screen Condition Cause Action
E-xx
E-1 through E-16, E-18
through E-99
Sensor not properly
connected to the sensor
cable.
Make sure the sensor is connected to the
sensor cable.
Sensor failure. Replace sensor.
Cable failure. • Replace sensor cable.
• Contact 3M Patient Warming technical
service.
E-xx
E-17 Sensor failure. Replace sensor.
Cable failure. • Replace sensor cable.
• Contact 3M Patient Warming technical
service.
Control unit failure. Discontinue use of the unit. Contact a
biomedical technician.
E-xx
E-100 and up Control unit failure. Discontinue use of the unit. Contact a
biomedical technician.