22 Specifications
English 202052A
Check the 3M
TM
SpotOn
TM
Temperature Monitoring System website to ensure you have the most recent version of this document. www.spotontemperature.com reorder #202052A
Electrical
characteristics
External power supply
100-240 VAC, 50-60 Hz
Output; 5 VDC
Class II, double insulated medical grade
Maximum heating power
2 W
Temperature
characteristics
Over-temperature cutoff
43°C - Skin thermistor reading
44.5°C - Heater thermistor reading
Performance
characteristics
Equilibration time
3 minutes
Time response
Approximately three minutes
Measurement locations
Patient’s lateral forehead above the orbital
ridge
Measurement range
25°C - 43°C
Measurement readout
Celsius or Fahrenheit
Leakage current
CF classication. Meets leakage current
requirements in accordance with UL 60601-1
and EN60601-1.
Environmental
conditions
Ambient temperature range
10°C to 40°C (50°F to 104°F)
Storage and transport temperature
range
-20°C to 60°C (-4°F to 140°F)
Store all components at room temperature
and in a dry place when not in use.
Operating humidity
10 to 75% RH, noncondensing
Atmospheric pressure range
80 kPa to 106 kPa (12 PSI to 15 PSI)
Oxygen environment
e sensor only is suitable for use in the
presence of an oxygen enriched environment.