3M eg4 Marine Instruments User Manual


 
Overview of running a study
Setting contrast on dosimeter
053-644, Rev H Page 36 eg4
36
Setting contrast on dosimeter
Contrast
1. Ensure the meter is Off.
2. Turn the meter on by pressing the on/off/enter button.
3. When the splash screen appears, press and hold the run/stop and on/off enter
buttons. (See figure 4-5 for splash screen).
4. The Setup menu appears on the Edge. Repeatedly press the arrow key until the Contrast
symbol is selected.
NOTE: Setup menu is used to reset the dosimeter, change the contrast, and return to the
main view.
Figure 4-7: Changing the contrast from the keypad
5. Next press the on/off enter button.
6. A number will appear on your display (i.e., 50). Press the up or down arrow to increase or
decrease the contrast of the Edge.
7. Press on/off enter button to accept the new contrast value.
8. To return to the measurement view screen, press the down arrow and press the
on/off/enter button.
Checkmark icon
Is the selection to
exit the setup menu
Contrast icon
Arrows and enter key
Indicates the valid keys
are selected