CALIBRATING THE INSTRUMENT DO 9704
Apply a voltage of 465.5 mVdc between pin 6 (GND) and pin 1 (OUT) of the DIN 8-pole connector
at input A.
Select step 11 with the PROG key; press the ENTER key, the display shows
Using the and keys, make the value shown on the display coincide with 196.0. Confirm by
pressing the ENTER key; P11 appears on the display; press PROG: P0 appears; press ENTER:
this quits the instrument calibration function.