Emerson Process Management 644H Marine Instruments User Manual


 
May 2014
16
Quick Start Guide
Product Certifications
Approved Manufacturing Locations
Rosemount Inc. - Chanhassen, Minnesota, USA
Rosemount Temperature GmbH – Germany
Emerson Process Management Asia Pacific – Singapore
European Directive Information
A copy of the EC Declaration of Conformity can be found at the end of the Quick Start Guide.
The most recent revision of the EC Declaration of Conformity can be found at
www.rosemount.com.
Ordinary Location Certification from FM Approvals
As standard, the transmitter has been examined and tested to determine that the design
meets the basic electrical, mechanical, and fire protection requirements by FM Approvals, a
nationally recognized test laboratory (NRTL) as accredited by the Federal Occupational
Safety and Health Administration (OSHA).
North America
E5 FM Explosionproof, Dust-Ignitionproof
Certificate: 3006278
Standards Used: FM Class 3600: 2011, FM Class 3615: 2006, FM Class 3616: 2011, FM
Class 3810: 2005, NEMA-250: 250: 2003, ANSI/IEC 60529: 2004
Markings: XP CL I, DIV 1, GP B, C, D; DIP CL II / III, GP E, F, G; T5(-50 °C T
a
+85 °C); Type
4X, IP66
I5 FM Intrinsic Safety and Nonincendive
Certificate: 3008880 [Headmount Fieldbus/Profibus, Railmount HART]
Standards Used: FM Class 3600: 1998, FM Class 3610: 2010, FM Class 3611: 2004, FM
Class 3810: 2005, NEMA - 250: 1991
Markings: IS CL I / II / III, DIV I, GP A, B, C, D, E, F, G; T4(-50 °C T
a
+60 °C); NI CL I, DIV 2,
GP A, B, C, D; T6(-50 °C T
a
+70 °C), T5(-50 °C T
a
+85 °C); when installed per
Rosemount drawing 00644-2075
Special Conditions for Safe Use (X):
1. When no enclosure option is selected, the Model 644 Temperature Transmitter shall be
installed in an enclosure meeting the requirements of ANSI/ISA S82.01 and S82.03 or
other applicable ordinary location standards
2. FM combination option code K5 is only applicable with Rosemount J5 Universal Head
(M20 x 1.5) or Rosemount J6 Universal Head (1/2-14 NPT) enclosure
3. Enclosure option has to be selected to maintain a Type 4X rating
Certificate: 3044581 [Headmount HART]
Standards Used: FM Class 3600: 2011, FM Class 3610: 2010, FM Class 3611: 2004, FM
Class 3810: 2005, ANSI/NEMA - 250: 1991; ANSI/IEC 60529: 2004; ANSI/ISA 60079-0:
2009; ANSI/ISA 60079-11: 2009
Markings: IS CL I / II / III, DIV I, GP A, B, C, D, E, F, G; IS Class I, Zone 0 A Ex ia IIC;
T4(-50 °C T
a
+80 °C); T5(-50 °C T
a
+40 °C); NI CL I, DIV 2, GP A, B, C, D;
T5(-50 °C T
a
+80 °C), T6(-50 °C T
a
+40 °C); when installed per Rosemount drawing
00644-2071; Type 4X; IP68