A SERVICE OF

logo

Integrated Cleaner Operation
A40-039 Rev 1 Page 16
W
ARNINGS
U
se only non-flammable, non-combustible water-based solutions. Don't ever use
alcohol, gasoline or flammable solutions. Do not use any solution that has a flash
point. Using flammable or combustible solutions may result in a fire and severe burns.
Always wear protective clothing; face shield and waterproof gloves and apron, when
handling cleaning solutions. Thermal and chemical burns may result from splashed or
spilled chemicals.
Always add chemicals to water. Never add water to chemicals. Adding water to
chemicals may result in violent reactions that splash hot chemicals and can cause
thermal or chemical burns.
Never pick up a solution filled tank. Splashing may cause thermal or chemical burns.
Do not overfill the tank. The tank may flood when large parts are inserted and cause
thermal or chemical burns.
Caution: Do not operate cleaner with the solution more than 4 inches below the top of the tank.
The heaters on the sides of the tank will be damaged if operated when not completely
covered by solution. The ultrasonics may be damaged if operated with no solution in the
tank.
4.7 Turning Power On
Press the POWER switch at the “1” to turn power on. The cleaner will run through a three second
self-test. The LED next to “SET DEGAS min” will light and the display will show the Degas set
time (in minutes).
4.8 Turning Power Off
Press the POWER switch at the “0” to turn power off. All functions will turn off and the displays
will be dark.
Note: Wait at least 10 seconds before turning POWER on after turning it off. If POWER is turned
on too soon, the cleaner will restart in the same configuration as it had when it was turned
off except that the control panel will be inoperative. Turn POWER off for 10 seconds to
return to normal operation with all functions off.
4.9 Degassing
Fresh solutions and solutions that have been idle for more than 24 hours contain dissolved gases
that reduce ultrasonic action. Operating the ultrasonics will remove these gases. Degas mode is
provided as a convenience to operate the ultrasonics for the time required to degas.
4.10 Setting Degas Time
Press the SELECT OPTION Switch until the LED next to “SET DEGAS min” lights. The
current Degas setpoint is displayed in Digital Number Display. The default time is 5 minutes.
Press CLEAR DISPLAY switch to set the time to 0.
Repeatedly press SET DISPLAY to set the desired Degas time between 01 and 99 minutes.