Integrated Cleaner Troubleshooting
A40-039 Rev 1 Page 21
Problem Cause What to do
Cleaner operates but
d
isplay does not
function.
Interrupted calibration se-
quence.
Malfunctioning timer board.
Press SET DISPLAY.
Call Branson Commercial
Products Department 800-731-
9162.
Decreased ultrasonic
a
ctivity.
Solution is not degassed. Make sure that tank was filled
with warm tap water plus
cleaning solution and has run
5-10 minutes.
S
olution is spent. Change solution.
S
olution level is incorrect for
load.
Adjust solution to operating
level line with load.
T
ank bottom is covered with
soil particles.
Empty, then clean tank with
warm water. Wipe with a
nonabrasive cloth.
U
sing deionized water in the
tank.
Deionized water does not
cavitate as actively as soapy
tap water.
Clogged drain. Clogged drain. U
nblock.