6. Important information
6. IMPORTANT INFORMATION
6.1 Safety info rmation
General recommendations
Read the safety and operating instructions before operating the equipment.
Retain safety and operating instructions for future reference.
Adhere to all warnings on the equipment and in the operating instructions manual.
Follow all instructions for operation and use.
Electrical shock
Image 6-1
Type of protection (Electrical)
Equipment with external power supply: Class I equipment
Degree of safety (Flammable anesthetic mixture)
Equipment not suitable for use in t he presence of a ammable anesthetic mixture with air or with oxygen
or nitrous oxide.
Non-patient care equipment
Equipment primarily for use in a health care facility that is intended for use where contact with a patient
is unlikely.
The equipment may not be used with life support equipme
nt.
Mission c ritical applications
We strongly recommend there is a replacement display immediately available in mission critical applica-
tions.
Power connection
Power requirements: The equipment must be powered using the delivered medical approved 24 VDC
power supply.
The medical approved DC power supply must be powered by the AC mains voltage (protective earth
terminal).
The equipment complies to the mentioned standards only when used with the supplied medical power
supply.
The equipment must be powered from a center-tapped electrical circuit when used in the USA at
voltages above 120 Volts.
The equipment is intended for continuous operation.
The equipment is powered from an external power supply for class I equipment. The installer is re-
sponsible for testing the equipment’s earth ground to verify that it meets the impedance requirements
for the given country regulatory requirements.
(451920610993)K5903021 MDSC-2226 01/03/2013