Emerson 1056 Marine Instruments User Manual


 
23
MODEL FCL-1056 SECTION 5.0
PROGRAMMING THE ANALYZER
5.3.3. Procedure – Configure Outputs.
1. Press MENU. The main menu screen appears. Move the cursor to
Program and press ENTER.
2. The cursor will be on Outputs. Press ENTER.
3. Choose Configure.
4. Choose Output 1 or Output 2.
5. The screen shows the present configuration. There are six items: Assign
(S1 is sensor 1, S2 is sensor 2), Range, Scale, Dampening, Fault
Mode, and Fault Value To display the fifth and sixth items, scroll to the
bottom of the screen and continue scrolling.
To make a change, move the cursor to the desired line and press ENTER.
A screen will appear in which the present setting can be edited. Press
ENTER to store the setting.
For an explanation of terms, see sections 5.3.1 and 5.3.2.
6. To return to the main display, press MENU then EXIT.
Program
Alarms
Measurement
Temperature
Outputs
Outputs
Range
Configure
Simulate
Configure
Output 1 Configure
Program
Output 2
Output 1
Output 1 Configure
S1 Meas
Range: 4-20 mA
Scale: Linear
Dampening 0 sec
Assign: