MODEL FCL-1056 SECTION 5.0
PROGRAMMING THE ANALYZER
5.9 RESETTING THE ANALYZER
5.9.1 Purpose
This section describes how to clear user-entered values and restore default settings. There are three resets:
1. Resetting to factory default values clears ALL user entered settings, including sensor and analog output
calibration, and returns ALL settings and calibration values to the factory defaults.
2. Resetting a sensor calibration to the default values clears user-entered calibration data for the selected
sensor but leaves all other user-entered data unaffected.
3. Resetting the analog output calibration clears only the user-entered analog output calibration. It leaves all
other user-entered settings unchanged.
1. Press MENU. The main menu screen appears. Move the cursor to
Program and press ENTER.
2. Scroll to the bottom of the screen and continue scrolling until Reset
Analyzer is highlighted. Press ENTER.
3. Choose whether to reset all user-entered values (Factory Defaults),
sensor calibration (Sensor Cal Only), or output calibration (Output Cal
Only). If you choose Sensor Cal Only or Output Cal Only a second
screen appears in which you can select which sensor or output
calibration to reset.
4. To return to the main display, press MENU then EXIT.
Reset Analyzer
Factory Defaults
Sensor Cal Only
Output Cal Only
Factory Defaults
Program
Temperature
Security
Diagnostics
Reset Analyzer
Reset Analyzer
5.9.2 Procedure – Resetting the Analyzer