Emerson 1056 Marine Instruments User Manual


 
24
MODEL FCL-1056 SECTION 5.0
PROGRAMMING THE ANALYZER
5.3.4. Procedure – Ranging Outputs.
1. Press MENU. The main menu screen appears. Move the cursor to
Program and press ENTER.
2. The cursor will be on Outputs. Press ENTER.
3. Choose Range.
4. Choose Output 1 or Output 2.
5. The screen shows the present settings for the outputs. O1 is output 1, O2
is output 2, S1 is sensor 1, and S2 is sensor 2.
To make a change, move the cursor to the desired line and press ENTER.
A screen will appear in which the present setting can be edited. Press
ENTER to store the setting.
For an explanation of terms, see sections 5.3.1 and 5.3.2.
6. To return to the main display, press MENU then EXIT.
Program
Alarms
Measurement
Temperature
Outputs
Outputs
Range
Configure
Simulate
Range
Output Configure
Program
Output 2
Output 1
Output Range
0.000 ppm
O1 S1 20mA: 10.00 ppm
O2 S1 4mA: 0.0C
O2 S1 20mA: 100.0C
O1 S1 4mA